Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI

Document Type: Tutorial
NI Supported: Yes
Publish Date: Jun 12, 2008


Feedback


Yes No

Related Categories

Related Links - Developer Zone

Related Links - Products and Services

Measuring Strain with Strain Gauges

949 ratings | 4.25 out of 5
Read in | Print

Overview

This tutorial is part of the National Instruments Measurement Fundamentals series. Each tutorial in this series will teach you a specific topic of common measurement applications by explaining theoretical concepts and providing practical examples.This tutorial introduces and explains the concepts and techniques of measuring strain with strain gauges.

You can also view an on demand webcast on strain gauge measurements.

For more information, return to the NI Measurement Fundamentals Main Page.

What Is Strain?

Strain is the amount of deformation of a body due to an applied force. More specifically, strain (e) is defined as the fractional change in length, as shown in Figure 1 below.



Figure 1. Definition of Strain



Strain can be positive (tensile) or negative (compressive). Although dimensionless, strain is sometimes expressed in units such as in./in. or mm/mm. In practice, the magnitude of measured strain is very small. Therefore, strain is often expressed as microstrain (me), which is e x 10-6.

When a bar is strained with a uniaxial force, as in Figure 1, a phenomenon known as Poisson Strain causes the girth of the bar, D, to contract in the transverse, or perpendicular, direction. The magnitude of this transverse contraction is a material property indicated by its Poisson's Ratio. The Poisson's Ratio n of a material is defined as the negative ratio of the strain in the transverse direction (perpendicular to the force) to the strain in the axial direction (parallel to the force), or n = eT/e. Poisson's Ratio for steel, for example, ranges from 0.25 to 0.3.

The Strain Gauge


While there are several methods of measuring strain, the most common is with a strain gauge, a device whose electrical resistance varies in proportion to the amount of strain in the device. The most widely used gauge is the bonded metallic strain gauge.

The metallic strain gauge consists of a very fine wire or, more commonly, metallic foil arranged in a grid pattern. The grid pattern maximizes the amount of metallic wire or foil subject to strain in the parallel direction (Figure 2). The cross sectional area of the grid is minimized to reduce the effect of shear strain and Poisson Strain. The grid is bonded to a thin backing, called the carrier, which is attached directly to the test specimen. Therefore, the strain experienced by the test specimen is transferred directly to the strain gauge, which responds with a linear change in electrical resistance. Strain gauges are available commercially with nominal resistance values from 30 to 3000 Ω, with 120, 350, and 1000 Ω being the most common values.


Figure 2. Bonded Metallic Strain Gauge


It is very important that the strain gauge be properly mounted onto the test specimen so that the strain is accurately transferred from the test specimen, through the adhesive and strain gauge backing, to the foil itself.

A fundamental parameter of the strain gauge is its sensitivity to strain, expressed quantitatively as the gauge factor (GF). Gauge factor is defined as the ratio of fractional change in electrical resistance to the fractional change in length (strain):



The Gauge Factor for metallic strain gauges is typically around 2.

Strain Gauge Measurement


In practice, the strain measurements rarely involve quantities larger than a few millistrain(e x 10-3). Therefore, to measure the strain requires accurate measurement of very small changes in resistance. For example, suppose a test specimen undergoes a strain of 500 me. A strain gauge with a gauge factor of 2 will exhibit a change in electrical resistance of only 2 (500 x 10-6) = 0.1%. For a 120 W gauge, this is a change of only 0.12 W.

To measure such small changes in resistance, strain gauges are almost always used in a bridge configuration with a voltage excitation source. The general Wheatstone bridge, illustrated below, consists of four resistive arms with an excitation voltage, VEX, that is applied across the bridge.


Figure 3. Wheatstone Bridge


The output voltage of the bridge, VO, will be equal to:



From this equation, it is apparent that when R1/R2 = R4/R3, the voltage output VO will be zero. Under these conditions, the bridge is said to be balanced. Any change in resistance in any arm of the bridge will result in a nonzero output voltage.

Therefore, if we replace R4 in Figure 3 with an active strain gauge, any changes in the strain gauge resistance will unbalance the bridge and produce a nonzero output voltage. If the nominal resistance of the strain gauge is designated as RG, then the strain-induced change in resistance, DR, can be expressed as DR = RG·GF·e. Assuming that R1 = R2 and R3 = RG, the bridge equation above can be rewritten to express VO/VEX as a function of strain (see Figure 4). Note the presence of the 1/(1+GF·e/2) term that indicates the nonlinearity of the quarter-bridge output with respect to strain.


Figure 4. Quarter-Bridge Circuit



Ideally, we would like the resistance of the strain gauge to change only in response to applied strain. However, strain gauge material, as well as the specimen material to which the gauge is applied, will also respond to changes in temperature. Strain gauge manufacturers attempt to minimize sensitivity to temperature by processing the gauge material to compensate for the thermal expansion of the specimen material for which the gauge is intended. While compensated gauges reduce the thermal sensitivity, they do not totally remove it.

By using two strain gauges in the bridge, the effect of temperature can be further minimized. For example, Figure 5 illustrates a strain gauge configuration where one gauge is active (RG + DR), and a second gauge is placed transverse to the applied strain. Therefore, the strain has little effect on the second gauge, called the dummy gauge. However, any changes in temperature will affect both gauges in the same way. Because the temperature changes are identical in the two gauges, the ratio of their resistance does not change, the voltage VO does not change, and the effects of the temperature change are minimized.


Figure 5. Use of Dummy Gauge to Eliminate Temperature Effects


The sensitivity of the bridge to strain can be doubled by making both gauges active in a half-bridge configuration. For example, Figure 6 illustrates a bending beam application with one bridge mounted in tension (RG + DR) and the other mounted in compression (RG - DR). This half-bridge configuration, whose circuit diagram is also illustrated in Figure 6, yields an output voltage that is linear and approximately doubles the output of the quarter-bridge circuit.

 


Figure 6. Half-Bridge Circuit


Finally, you can further increase the sensitivity of the circuit by making all four of the arms of the bridge active strain gauges in a full-bridge configuration. The full-bridge circuit is shown in Figure 7.

 


Figure 7. Full-Bridge Circuit


The equations given here for the Wheatstone bridge circuits assume an initially balanced bridge that generates zero output when no strain is applied. In practice however, resistance tolerances and strain induced by gauge application will generate some initial offset voltage. This initial offset voltage is typically handled in two ways. First, you can use a special offset-nulling, or balancing, circuit to adjust the resistance in the bridge to rebalance the bridge to zero output. Alternatively, you can measure the initial unstrained output of the circuit and compensate in software.

The equations given above for quarter, half, and full-bridge strain gauge configurations assume that the lead wire resistance is negligible. While ignoring the lead resistances may be beneficial to understanding the basics of strain gauge measurements, doing so in practice can be a major source of error. For example, consider the 2-wire connection of a strain gauge shown in Figure 8a. Suppose each lead wire connected to the strain gauge is 15 m long with lead resistance RL equal to 1 W. Therefore, the lead resistance adds 2 W of resistance to that arm of the bridge. Besides adding an offset error, the lead resistance also desensitizes the output of the bridge.

You can compensate for this error by measuring the lead resistance RL and accounting for it in the strain calculations. However, a more difficult problem arises from changes in the lead resistance due to temperature fluctuations. Given typical temperature coefficients for copper wire, a slight change in temperature can generate a measurement error of several me.

Using a 3-wire connection can eliminate the effects of variable lead wire resistance because the lead resistances affect adjacent legs of the bridge. As seen in Figure 8b, changes in lead wire resistance, RL2, do not change the ratio of the bridge legs R3 and RG. Therefore, any changes in resistance due to temperature cancel each other.



Figure 8. 2-Wire and 3-Wire Connections of Quarter-Bridge Circuit

Signal Conditioning for Strain Gauges

Strain gauge measurement involves sensing extremely small changes in resistance. Therefore, proper selection and use of the bridge, signal conditioning, wiring, and data acquisition components are required for reliable measurements. To ensure accurate strain measurements, it is important to consider the following:

  • Bridge completion
  • Excitation
  • Remote sensing
  • Amplification
  • Filtering
  • Offset
  • Shunt calibration


Bridge Completion – Unless you are using a full-bridge strain gauge sensor with four active gauges, you will need to complete the bridge with reference resistors. Therefore, strain gauge signal conditioners typically provide half-bridge completion networks consisting of high-precision reference resistors. Figure 9 shows the wiring of a half-bridge strain gauge circuit to a conditioner with completion resistors R1 and R2.


Figure 9. Connection of Half-Bridge Strain Gauge Circuit


Excitation – Strain gauge signal conditioners typically provide a constant voltage source to power the bridge. While there is no standard voltage level that is recognized industry wide, excitation voltage levels of around 3 and 10 V are common. While a higher excitation voltage generates a proportionately higher output voltage, the higher voltage can also cause larger errors because of self-heating.

Remote Sensing – If the strain gauge circuit is located a distance away from the signal conditioner and excitation source, a possible source of error is voltage drop caused by resistance in the wires connecting the excitation voltage to the bridge. Therefore, some signal conditioners include a feature called remote sensing to compensate for this error. Remote sense wires are connected to the point where the excitation voltage wires connect to the bridge circuit. The extra sense wires serve to regulate the excitation supply through negative feedback amplifiers to compensate for lead losses and deliver the needed voltage at the bridge.

Amplification – The output of strain gauges and bridges is relatively small. In practice, most strain gauge bridges and strain-based transducers will output less than 10 mV/V (10 mV of output per volt of excitation voltage). With 10 V excitation, the output signal will be 100 mV. Therefore, strain gauge signal conditioners usually include amplifiers to boost the signal level to increase measurement resolution and improve signal-to-noise ratios.

Filtering – Strain gauges are often located in electrically noisy environments. It is therefore essential to be able to eliminate noise that can couple to strain gauges. Lowpass filters, when used in conjunction with strain gauges, can remove high-frequency noise prevalent in most environmental settings.

Offset Nulling – When a bridge is installed, it is very unlikely that the bridge will output exactly zero volts when no strain is applied. Slight variations in resistance among the bridge arms and lead resistance will generate some nonzero initial offset voltage. Offset nulling can be performed by either hardware or software:

1. Software Compensation – With this method, you take an initial measurement before strain input is applied, and use this offset to compensate subsequent measurements. This method is simple, fast, and requires no manual adjustments. The disadvantage of the software compensation method is that the offset of the bridge is not removed. If the offset is large enough, it limits the amplifier gain you can apply to the output voltage, thus limiting the dynamic range of the measurement.

2. Offset-Nulling Circuit – The second balancing method uses an adjustable resistance, a potentiometer, to physically adjust the output of the bridge to zero. By varying the resistance of potentiometer, you can control the level of the bridge output and set the initial output to zero volts.

Shunt Calibration – The normal procedure to verify the output of a strain gauge measurement system relative to some predetermined mechanical input or strain is called shunt calibration. Shunt calibration involves simulating the input of strain by changing the resistance of an arm in the bridge by some known amount. This is accomplished by shunting, or connecting, a large resistor of known value across one arm of the bridge, creating a known DR. The output of the bridge can then be measured and compared to the expected voltage value. The results are used to correct span errors in the entire measurement path, or to simply verify general operation to gain confidence in the setup.

DAQ Systems for Strain Gauge Measurements


Using SCXI with Strain Gauges

National Instruments SCXI is a signal conditioning system for PC-based instrumentation applications. An SCXI system consists of a shielded chassis that houses a combination of signal conditioning input and output modules, which perform a variety of signal conditioning functions. You can connect many different types of sensors, including strain gauges, directly to SCXI modules. The SCXI system operates as a front-end signal conditioning system for PC plug-in data acquisition (DAQ) devices (PCI and PCMCIA) or PXI DAQ modules.


Figure 10. SCXI Signal Conditioning System

The SCXI-1520 is an 8-channel universal strain-gauge input module that offers a variety of features for strain measurements. With this single module, signals from strain, force, torque, and pressure sensors can be easily read. The SCXI-1520 also offers a programmable amplifier and programmable 4-pole Butterworth filter on each channel, and simultaneous sampling with track-and-hold circuitry. In addition, the SCXI-1520 system offers a half-bridge completion resistor network in the module and a socketed 350 W quarter-bridge completion resistor. Table 1 summarizes some additional features of the SCXI-1520 that relate to strain gauge measurements.


Table 1. SCXI-1520 Features for Strain Gauges
Number of channels 8
Multiplexer scan rate Up to 333 kS/s1
Amplifier gain 1 to 1000
Excitation voltage source 0.0 to 10.0 V in 0.635 V increments
Excitation current drive 29 mA throughout excitation voltage range
Half-bridge completion Yes
Offset nulling Yes
Shunt calibration Yes
Remote excitation sensing Yes
1 Multiplexer scan rate depends on DAQ device.


Recommended Starter Kit for Strain Gauge SCXI DAQ System:

1. PCI-6052 DAQ board
2. SCXI-1000 chassis
3. SCXI-1349 cable assembly
4. SCXI-1520 with SCXI-1314 terminal block
5. Refer to ni.com/sensors for recommended sensor vendors


Using SCC with Strain Gauge Measurements

National Instruments SCC provides portable, modular signal conditioning for DAQ systems. SCC can condition a variety of analog I/O and digital I/O signals. SCC DAQ systems include an SC-2345 Series shielded carrier, SCC modules, a cable, and a DAQ device. Figure 11 below illustrates an SC-2345 carrier with SCC modules.



Figure 11. SC-2345 with SCC Modules



SCC-SG Series modules can be used for conditioning quarter, half, and full-bridge strain gauges. Each module has two strain gauge input channels, offset nulling circuitry for each channel, and a 2.5 V excitation circuit. Each input channel includes an instrumentation amplifier with differential inputs and a fixed gain of 100. The output of each amplifier is filtered and buffered to prevent settling time delays. The SCC-SG01 works with 120 W quarter-bridge strain gauges, and the SCC-SG02 works with 350 W quarter-bridge strain gauges. The SCC-SG03 works with half-bridge strain gauges and the SCC-SG04 works with full-bridge strain gauges. Figure 12 illustrates the block diagram of the SCC-SG01/01 modules.


Figure 12. Block Diagram of the SCC-SG01, SCC-SG02 Quarter-Bridge Modules



The SCC-SG11 is a dual-channel shunt calibration module for use with the SCC-SG Series Modules. Each channel includes two terminals for wiring a switched 301 kW, 1%, 1/4 W resistor across both channels of a module by writing a logic high to the digital line controlling the SCC-SG11. You can disable shunt calibration by writing a logic low to the same digital line. Figure 14 illustrates an example of using the SCC-SG11 shunt calibration module with the SCC-SG02 module.


Figure 13. Using the SCC-SG11 Shunt Calibration Module with the SCC-SG02



Recommended Starter Kit for Strain Gauge SCC DAQ System:

1. PCI-6052 DAQ board
2. SC-2345 module carrier
3. SCC-SG01/02 (quarter-bridge), SCC-SG03 (half-bridge), or SCC-SG04(full-bridge)
4. SCC-SG11 (shunt calibration)
5. Refer to ni.com/sensors for recommended strain gauges and full-bridge sensors.

Relevant NI Products


Customers interested in this topic were also interested in the following NI products:

For more tutorials, return to the NI Measurement Fundamentals Main Page.


949 ratings | 4.25 out of 5
Read in | Print

Reader Comments | Submit a comment »

very informative article
very helpful article on priliminary learning about strain guages. cheers!! great work
- harminder saini, Student, brunel university. bs07hss@brunel.ac.uk - Jul 8, 2008

Awesome!
- Apr 16, 2008

An extremely helpful and self explanatory article
- ahsan dar, PN Engineering College,NUST. ahsandar@hotmail.com - Dec 18, 2007

most useful information
This is most useful information for me and my student because Iam teaching manufacturing automation subject
- SONAWANE V. Y. , AISSMS COLLEGE OF ENGG NEAR RTO KENEDY ROAD PUNE-1. sonawanevy@yahoo.co.in - Oct 26, 2007

i have a 3 hour exam in the morning. superb refresher. Cheers
- May 13, 2007

very useful
Thank you for providing great information about strain guage sensor which helped me a lot in designing this sensor practically.
- vamshidhar.smt@gmail.com - Apr 5, 2007

Thank You Much. I can now start my tech paper. Just what I needed to get started
- Starving Student - Apr 3, 2007

Need numerical results
Information provided is very good and useful.Come up with practical results eg:what is the range of usability of the instrument,sensitivity etc
- vamshidhar s, IIT Kharagpur. vamshidhar.smt@gmail.com - Mar 18, 2007

this information about strain gauge is not enough . this is just copy from book. i thing u have done it practically.
- jassi, infotech. jaspreetd@rediffmail.com - Mar 13, 2007

Comprehensive article
Thank you for this elaborative content. It gives me a kick start for an indepth strain measurement activities.
- Demissie Solomon, Ethiopian Airlines. solomond@ethiopianairlines.com - Dec 20, 2006

Self explainatory
From basic to higher end , it is very technical.
- awachar@yahoo.com - Dec 5, 2006

Diverse, robust, and thorough. Easy to read and practical.
- Nov 2, 2006

Brilliant!
Very basic, yet a comprehensive article. 5/5
- Usman Y. Bhatti, All Sensors. usmanyasin@gmail.com - Oct 10, 2006

What a great work!!
Thanks for your information!!it's very useful for me.thanks!!
- Ajmal Khan, International Islamic University Malaysia,Kuala Lumpur. ir_mal001@yahoo.com - Sep 30, 2006

thaaaaaaaaanx
- Charanpreet Singh. rahelcharanpreet@raediffmail.com - Sep 24, 2006

excellent
wonderfull explanation.covers basics very well
- ritula thakur, Faculty in NITTTR,chandigarh. ritula_212yahoo.com - Jun 5, 2006

Keep it up!
Great job, helpful with school research project
- Kaycee, Coventry University, UK. masterc34@yahoo.co.uk - May 4, 2006

helpful but needs more examples on how to calculate resistance values.
- Louis, McGill U. - Apr 20, 2006

Perfect theory about concepts of electrical strain gauge
It is a very nice effort covering the basic concepts of electrical strain gauge. Your effort became very useful for me. Its a perfect tutorial.
- Haris Aslam, University of Engineering and Technology Lahore Pakistan. haris_uetian@yahoo.com - Apr 14, 2006

i think ...it is the best document on gauges available on the internet
- brij, iit kharagpur. - Apr 10, 2006

Informative on an introductory level
- Damien Lea, Aerospace Engineering Ltd. damienlea@yahoo.com - Mar 25, 2006

very helpful...
- gilbert, Copenhagen University. - Mar 17, 2006

strain measurement fundas
The article gives requisite optimum and information on the measurement system that is practical in application without flambuayant add ons.
- v.k.khanna, Intertech Sales Corporation. intertec@bol.net.in - Mar 7, 2006

- NICE & GOOD - Mar 5, 2006

Man That was gr8.. we need this stuff to do our Main Project.. a teleremote controlled robotic hand.. THANX!!!
- Anoop Kumar, NSS College of engineering. - Feb 21, 2006

great
- Ubaier, Coventry University, UK. - Jan 13, 2006

Clear and useful
- Kate - Dec 16, 2005

Thankyou for this resourse.
- Oct 31, 2005

very good!! thanks for the valuable information!! =)
- teo - Aug 31, 2005

really informative.good work
- sanju thomas, vellore institute of technology. sanjuthom@gmail.com - Jul 13, 2005

excellent
This document is very much helpful for me to do my project work, thanks to NI
- anbalagan, PSG College of Technology,tamilnadu,india. anbuxyz@yahoo.co.in - May 12, 2005

basics covered well
basics covred with visulaization help.Good for preliminary info.
- Ravi, IIT,Kharagpur,. thebroadmind@rediffmal.com - May 3, 2005

Really Good!
Very Well Written. Keep It Up
- Kiran - May 2, 2005

doing M.tech from IIT kharagpur
very good site giving useful information... thank you
- rajani kanth gandla, student. grajanikanth@yahoo.com - Apr 21, 2005

Great help with my primary school project
Great help with my primary school project - thanks!
- The child prodigy - Mar 8, 2005

How very spiffing.
I really enjoyed this "web page" on the Intranet. I am looking into how 'phones work and this page was really useful to me. Thank- you so much for creating such a useful site.
- Jim - Jan 20, 2005

Seems brilliant. Just what I need.
- Jan 18, 2005

Good work on theory. However, more description on how to configure with your system would be helpful i.e SCXI 1314 wiring diagram.
- Aug 18, 2004

Clear and concise. Could benefit from coverage of practical considerations - how to measure the right strain
- Aug 18, 2004

A strain free coverage of the subject.
- Moses Zulfikar, Domino Printing UK. mzulfikar@waitrose.com - Aug 5, 2004

Great Help for my project!!!
- Edwin - Jun 7, 2004

Great summary; please add cFP DAQ!
This great summary on signal conditioning for strain gauges will be complete when You add FP and cFP to the already covered SCXI and SCC DAQ options!
- Richard Larker, Indexator AB. Richard.Larker@indexator.se - May 24, 2004

Thanks
Great help with the 2004 AS Physics Practical Assesment 19th May
- Xylomn. xylomn@hotmail.com - May 18, 2004

great help with physics AS planning exercise
supplied a lot of the information I needed to find for my AS Physics planning exercise. Thanks a lot
- Sarah Osborne-White. miniarmygirl@hotmail.com - May 16, 2004

great help with school project
not too wordy. told me everything I needed to know. thank you
- Iain Stenson. iaianstenson22@hotmail.com - May 10, 2004

Thank you
Just done a HNC project on this and i'd like to say this cleared up several issue and gave more background to my write up
- Anon - May 9, 2004

Very clear and helpful
- Graham - May 8, 2004

Very informative, but...
Very informative of the electrical and measurement aspects, although being number 1 google result, I think there'll be a lot of quarter bridges used. This being for A-Level physicss coursework. The main downside is it doesn't advise me how to fix a strain guage to a surface. With Super Glue?
- Chiral, member.lycos.co.uk/chiralutd/phpBB2/index.php. chiralutd@yahoo.co.uk - Mar 23, 2004

A thorough approch
A thorough scientific and emperical approch. Keep it up !!! Kind regards Nicholas Lekatsas-Lucas Senior electronic R/D Eng.
- Nicholas Lekatsas-Lucas, ICE/HT-FORTH. nicklek@iceht.forth.gr - Nov 15, 2002

excellent coverage of the electronic theory behind strain gauge measurement
Question: I'm searching for a practical, unbiased performance comparison between tension measurement devices that use silicon-type strain gauges versus those that use foil-type strain gauges. What are the advantages of one technology over the other? Could you point me toward any research that's been done?
- Mark Breen, Dover Flexo Electronics. mbreen@dfe.com - Oct 8, 2002

good but improvements possible
This is good but theres room for info about its applications and advantages/comparisons to other methods
- Sep 22, 2002

Very clear and concise - good job
- Jul 9, 2002

excellent presentation helping to understand the process of selecting the right meter!
- mstraka@scientech.com - Apr 17, 2002

hi, im an electromechanical engineer student, the information given here was great, i just think that you have to translate it, at 2 or 3 languajes more.. i did in spanish =). tanx
- crdh@pa.inter.net - Aug 11, 2001

 

Legal
This tutorial (this "tutorial") was developed by National Instruments ("NI"). Although technical support of this tutorial may be made available by National Instruments, the content in this tutorial may not be completely tested and verified, and NI does not guarantee its quality in any way or that NI will continue to support this content with each new revision of related products and drivers. THIS TUTORIAL IS PROVIDED "AS IS" WITHOUT WARRANTY OF ANY KIND AND SUBJECT TO CERTAIN RESTRICTIONS AS MORE SPECIFICALLY SET FORTH IN NI.COM'S TERMS OF USE (http://ni.com/legal/termsofuse/unitedstates/us/).