NI 5122/5124 Video Triggering
Overview
Return to Fundamentals of High-Speed Digitizers
The NI 5122/5124 includes a mode for triggering on NTSC, PAL, and SECAM format video signals. The NI 5122/5124 supports triggering on any line, a specific line number, or a specific field. The NI 5122/5124 supports negative trigger polarity in video mode, which means that the synchronization pulses must go to a negative voltage with respect to the back porch level.
| Note Using a vertical range that is too high may cause the digitizer to not trigger correctly on the video signal. |
Video Signal DC Restoration
NI 5122/5124 digitizers support DC restoration, which adjusts the back porch level of the video signal to 0 V. DC restore is performed on the digital data, rather than with an analog clamping circuit. When using DC restore, set the input voltage range to approximately twice the expected signal amplitude to prevent signal clipping as the amplitude and offset of the video signal change. DC restore measures the voltage of the back porch during the interval between the end of the color burst and before the start of the active line, then subtracts this value from the signal for the remainder of the active video line. Excess noise on the back porch portion of the video signal may cause inconsistent level restoration between lines.
Signal Conditioning
To provide proper termination to the video signal, use a 75 Ω terminator on the input of the digitizer in 1 MΩ input mode.
Copy Protection System Implications
Many video signal sources, particularly DVD players, produce a copy protection signal. DVDs can be encoded to enable this function, which inhibits copying the signal by injecting pulses that violate video signal standards, and disrupts input circuitry on VCRs. These signals often include false synchronization pulses and signals that exceed the voltage range of a standard signal. Any false pulses may cause erroneous line triggering on the NI 5122/5124. To avoid this error in your application, try one of the following solutions:
- Deactivate or remove the copy protection signal from the signal source.
- Recover triggering information in software by taking an untriggered waveform with the NI 5122/5124 and searching for valid trigger points in the waveform.
Fundamentals of High-Speed Digitizers
Video Signal Fundamentals
Reader Comments | Submit a comment »
Missing critical quality information
Absolutely any reference material concerning
using the 5122 to capture NTSC/PAL/SECAM
video *must* warn the potential user/buyer
that the 5122 does not have a 75 ohm
impedance selection. This makes it
completely unsuitable for video measurements.
- Richard Myers, Bosch Security System. richard.myers@us.bosch.com - Aug 12, 2005
Legal
This tutorial (this "tutorial") was developed by National Instruments ("NI"). Although technical support of this tutorial may be made available by National Instruments, the content in this tutorial may not be completely tested and verified, and NI does not guarantee its quality in any way or that NI will continue to support this content with each new revision of related products and drivers. THIS TUTORIAL IS PROVIDED "AS IS" WITHOUT WARRANTY OF ANY KIND AND SUBJECT TO CERTAIN RESTRICTIONS AS MORE SPECIFICALLY SET FORTH IN NI.COM'S TERMS OF USE (http://ni.com/legal/termsofuse/unitedstates/us/).