NI-5411 Arbitrary Waveform Generator - Single Trigger Mode
through the entire staging list. Only one trigger is required to start the
waveform generation.
You can use single trigger mode with both the ARB and DDS waveform
generation modes as follows:
ARB mode
Figure 1 uses the stages 1, 2, and 3 shown in Figure 2-4
to illustrate a single trigger mode of operation for ARB waveform
generation mode. After the NI 5411/5431 receives a trigger, the
waveform generation starts at the first stage and continues through the
last stage. The last stage is generated repeatedly until you stop the
waveform generation.

Figure 1 Single Trigger Mode for ARB Mode
Note: You can settle to a predefined state by making the last stage emulate that state.
DDS mode
After the NI 5411/5431 receives a trigger, the waveform
generation starts at the first stage and continues through the last stage.
The last stage is generated repeatedly until you stop the waveform
generation. Figure 2 illustrates a single trigger mode of operation for
DDS waveform generation mode.

Figure 2. Single Trigger Mode for DDS Mode
For example, assume that one cycle of a sine wave is stored in the DDS
lookup memory. For stage 1, f1 specifies the sine frequency to be
generated for time, T2 for stage 2, and so on. If there are
four stages in the staging list, f4 will be generated continuously until
the waveform generation is stopped.
Reader Comments | Submit a comment »
Legal
This tutorial (this "tutorial") was developed by National Instruments ("NI"). Although technical support of this tutorial may be made available by National Instruments, the content in this tutorial may not be completely tested and verified, and NI does not guarantee its quality in any way or that NI will continue to support this content with each new revision of related products and drivers. THIS TUTORIAL IS PROVIDED "AS IS" WITHOUT WARRANTY OF ANY KIND AND SUBJECT TO CERTAIN RESTRICTIONS AS MORE SPECIFICALLY SET FORTH IN NI.COM'S TERMS OF USE (http://ni.com/legal/termsofuse/unitedstates/us/).