Features Supported by Traditional NI-DAQ (Legacy) Devices

NI High-Speed Digitizers Help (NI-SCOPE 16.1)

Edition Date: January 2017

Part Number: 370592AB-01

View Product Info

DOWNLOAD (Windows Only)


NI-SCOPE 18.7 (English | Japanese)

The tables in this topic show the features supported by the NI 5102/5112/5620/5621. These digitizers are also known as Traditional NI-DAQ (Legacy) devices because they are configurable under Traditional NI-DAQ in Measurement & Automation Explorer (MAX).

Categories

Input Configuration

Timing

Triggering

Clocking

Acquisition

Synchronization Triggers

Miscellaneous


Input Configuration

NI 5102 NI 5112 NI 5620 NI 5621
Maximum Input Range ±5 V ±25 V ±1 V (±10 dBm) ±1 V (±10 dBm)
DC Offset N Y N N
Input Impedance 1 MΩ 50 Ω,
1 MΩ
50 Ω 50 Ω
Input Coupling AC, DC AC, DC AC only DC only
3 dB Bandwidth 0 MHz to 15 MHz 0 MHz to 100 MHz
or 0 MHz to 20 MHz
(analog filter)
10 kHz to 36 MHz 0 MHz to 36 MHz

Timing

NI 5102 NI 5112 NI 5620 NI 5621
Maximum Real-Time Sample Rate 20 MS/s 100 MS/s 64 MS/s 64 MS/s
Maximum RIS Sample Rate 1 GS/s 2.5 GS/s
Time Stamps Relative Absolute, Relative Absolute, Relative Absolute, Relative

Triggering

NI 5102 NI 5112 NI 5620 NI 5621
Reference (Stop) Trigger Types Immediate, Analog Edge, Digital Edge, Hysteresis, Window Immediate, Analog Edge, Digital Edge, Hysteresis, Window, Software Immediate, Digital Edge, Software, Analog Edge, Analog Window, Analog Hysteresis Immediate, Analog Edge, Digital Edge, Hysteresis, Window, Software
Reference (Stop) Trigger Sources CH 0,
CH 1,
TRIG,
PFI<1..2>,
RTSI<0..6>,
PXI STAR
CH 0, CH 1, TRIG, PFI<1..2>, RTSI<0..6>, PXI STAR PFI 1,
RTSI<0..6>,
PXI STAR
CH 0,
PFI<1..2>,
RTSI<0..6>
Trigger Coupling (External TRIG Input) AC, DC AC, DC
Trigger Coupling (Analog Input Channel) Same as input channel AC, DC Same as input channel Same as input channel
Trigger Holdoff From end of acquisition to trigger From trigger to first point of next record
Trigger Delay N Y N N
External TRIG Input Impedance 1 MΩ 1 MΩ, 50 Ω

Clocking

NI 5102 NI 5112 NI 5620 NI 5621
Reference Clock Input Sources PFI<1..2>,
RTSI Clock (RTSI 7)
PFI<1..2>,
RTSI Clock (RTSI 7),
PXI Clock 10
REF CLK IN,
PXI CLK 10
REF CLK IN,
PXI CLK 10
Reference Clock Frequency 20 MHz 10 MHz 10 MHz 10 MHz
Reference Clock Outputs PFI<1..2>,
RTSI Clock (RTSI 7)
PFI<1..2>,
RTSI Clock (RTSI 7)
PCI Internal Reference Clock Outputs PFI<1..2>,
RTSI Clock (RTSI 7)
PFI<1..2>,
RTSI Clock
(RTSI 7)
External Sample Clock Sources PFI<1..2>,
RTSI<0..6>
External Sample Clock Frequency 1 kHz to 20 MHz
Sample Clock Outputs PFI<1..2>,
RTSI<0..6>

Acquisition

NI 5102 NI 5112 NI 5620 NI 5621
Acquisition Modes Normal Normal Normal, DDC Normal, DDC
Multiple Records N Y Y Y
Continuous Acquisition N Y Y Y

Synchronization Triggers

NI 5102 NI 5112 NI 5620 NI 5621
Arm Reference Trigger Sources
Record Advance Trigger Sources PFI<0..1>,
RTSI<0..7>,
PXI STAR
Record Arm Trigger Sources PFI<1..2>,
RTSI<0..6>
Start Trigger (Acquisition Arm) Sources PFI<1..2>,
RTSI<0..6>,
PXI STAR
Reference (Stop) Trigger Outputs PFI<1..2>,
RTSI<0..6>
PFI<1..2>,
RTSI<0..6>,
PXI STAR
(PXI Slot 2 only)
PFI 1,
RTSI<0..6>
PFI 1,
RTSI<0..6>
End of Record Outputs PFI<1..2>,
RTSI<0..6>
End of Acquisition Outputs
Start Trigger Outputs PFI<1..2>,
RTSI<0..6>
Master–Slave Trigger Delay Correction N Y N N

Miscellaneous

NI 5102 NI 5112 NI 5620 NI 5621
Self-Calibration RIS Timing Only Y N N
Probe Compensation PFI 1 (1 kHz) PFI 1 (1 kHz)
LabVIEW Real-Time Support N N N N
Memory per Channel 663,000 16 or 32 MB 16 or 32 MS (not using DDC),
8 or 16 MS (using DDC)
32 MS (not using DDC),
16 MS (using DDC)
Sample Width 8 bits 8 bits 16 bits 16 bits


Note  Refer to the specifications document that shipped with your device for more information.