NI 5132/5133 digitizers support single-record acquisitions only. The following timing diagram illustrates how NI 51352/5133 digitizers react to the user-configurable input triggers during acquisition.

Each state prefixed by Wait for is a state in which an input trigger can be configured. This trigger tells the digitizer when to transition out of that particular state. The hardware is only sensitive to a particular trigger when in that trigger's appropriate Wait for state.
![]() |
Note The trigger signals in the timing diagram assume active high level triggers. |
The exportable event (End of Acquisition) is also shown in the timing diagram. The End of Acquisition Event asserts when the record has been completed, but it does not assert if the acquisition is aborted or times out.