NIDCPOWER_ATTR_SAMPLES_TO_AVERAGE

NI DC Power Supplies and SMUs Help (NI-DCPower)

Edition Date: January 2018

Part Number: 370736U-01

View Product Info

DOWNLOAD (Windows Only)


NI-DCPower (English | Japanese)

Specific Attribute

Numeric Value Data
type
Access Applies to Coercion High-Level Functions
1150003 ViInt32 R/W Channel None None

Description

Specifies the number of samples to average when you take a measurement.

Increasing the number of samples to average decreases measurement noise but increases the time required to take a measurement.

Refer to the NI PXI-4110, NI PXI-4130, NI PXI-4132, or NI PXIe-4154 averaging topic for optional property settings to improve immunity to certain noise types.

For information about improving noise immunity for NI-DCPower devices that support DC noise rejection, refer to Measurement Noise Rejection.

Device Range of Samples to Average
NI PXI-4110 and NI PXI-4130 1 to 511 samples
NI PXI-4132 1 to 127 samples
NI PXIe-4112/4113 1
PXIe-4135 1
NI PXIe-4136/4137 1
NI PXIe-4138/4139 1
NI PXIe-4140/4141/4142/4143/4144/4145 1
NI PXIe-4154 1 to 65,535 samples
PXIe-4162/4163 1

Default Value:

Refer to Supported Attributes by Device for the default value by device.