niDCPower Configure Output Enabled VI

NI DC Power Supplies and SMUs Help (NI-DCPower)

Edition Date: January 2018

Part Number: 370736U-01

View Product Info

DOWNLOAD (Windows Only)


NI-DCPower (English | Japanese)
Installed With: NI-DCPower

Enables or disables generation on the specified channel(s).

Depending on the selected output function, the voltage level, current level, or output resistance must be set in addition to enabling the output to generate the desired level.

For more information about configuring the output level, refer to the niDCPower Configure Output Function VI.

Note  If the device is in the Uncommitted state, enabling the output does not take effect until you call the niDCPower Initiate VI.

niDCPower Configure Output Enabled

/images/reference/en-XX/help/370736U-01/cio.gif

instrument handle identifies a particular instrument session. instrument handle is obtained from the niDCPower Initialize With Channels VI.

/images/reference/en-XX/help/370736U-01/cstr.gif

channel name specifies to which output channel(s) to apply this configuration value. Specify multiple channels by using a channel list or a channel range. A channel list is a comma (,) separated sequence of channel names (for example, 0,2 specifies channels 0 and 2). A channel range is a lower bound channel followed by a hyphen (-) or colon (:) followed by an upper bound channel (for example, 0-2 specifies channels 0, 1, and 2). In the Running state, multiple output channel configurations are performed sequentially based on the order specified in this parameter.

/images/reference/en-XX/help/370736U-01/cbool.gif

output enabled specifies whether the output channel is enabled or disabled.

true Enables generation on the specified output channel(s).
false Disables generation on the specified output channel(s). This parameter has no effect on the output disconnect relay. To toggle the relay, use the Output Connected property.

The default value is true if you use the niDCPower Initialize With Channels VI to open the session. Otherwise the default value is false, including when you use a calibration session or the deprecated programming model.

/images/reference/en-XX/help/370736U-01/cerrcodeclst.gif

error in (no error) describes error conditions that occur before this node runs. This input provides standard error in functionality.

/images/reference/en-XX/help/370736U-01/cbool.gif

status is TRUE (X) if an error occurred before this VI ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI ran. The default is FALSE.

/images/reference/en-XX/help/370736U-01/ci32.gif

code is the error or warning code. The default is 0. If status is TRUE, code is a non-zero error code. If status is FALSE, code is 0 or a warning code.

/images/reference/en-XX/help/370736U-01/cstr.gif

source describes the origin of the error or warning and is, in most cases, the name of the VI that produced the error or warning. The default is an empty string.

/images/reference/en-XX/help/370736U-01/iio.gif

instrument handle out passes the handle used to identify the session in all subsequent NI-DCPower VI calls.

/images/reference/en-XX/help/370736U-01/ierrcodeclst.gif

error out contains error information. This output provides standard error out functionality.

/images/reference/en-XX/help/370736U-01/ibool.gif

status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.

/images/reference/en-XX/help/370736U-01/ii32.gif

code is the error or warning code. If status is TRUE, code is a non-zero error code. If status is FALSE, code is 0 or a warning code.

/images/reference/en-XX/help/370736U-01/istr.gif

source describes the origin of the error or warning and is, in most cases, the name of the VI that produced the error or warning.

WAS THIS ARTICLE HELPFUL?

Not Helpful