niDCPower Configure Voltage Level VI

NI DC Power Supplies and SMUs Help (NI-DCPower)

Edition Date: January 2018

Part Number: 370736U-01

View Product Info

DOWNLOAD (Windows Only)


NI-DCPower (English | Japanese)
Installed With: NI-DCPower

Configures the voltage level that the device attempts to generate for the specified channel(s). The channel must be enabled for the specified voltage level to take effect.

Refer to the niDCPower Configure Output Enabled VI for more information about enabling the output channel. The voltage level setting is applicable only if the channel is set to the DC Voltage output function using the niDCPower Configure Output Function VI.

The device actively regulates the voltage at the specified level unless doing so causes a current drop greater than the current limit across the channels' output terminals.

Related topics:

Constant Voltage Mode

niDCPower Configure Voltage Level

/images/reference/en-XX/help/370736U-01/cio.gif

instrument handle identifies a particular instrument session. instrument handle is obtained from the niDCPower Initialize With Channels VI.

/images/reference/en-XX/help/370736U-01/cstr.gif

channel name specifies to which output channel(s) to apply this configuration value. Specify multiple channels by using a channel list or a channel range. A channel list is a comma (,) separated sequence of channel names (for example, 0,2 specifies channels 0 and 2). A channel range is a lower bound channel followed by a hyphen (-) or colon (:) followed by an upper bound channel (for example, 0-2 specifies channels 0, 1, and 2). In the Running state, multiple output channel configurations are performed sequentially based on the order specified in this parameter.

/images/reference/en-XX/help/370736U-01/cdbl.gif

voltage level specifies the voltage level, in volts, for the output channel generation.

The valid values for this parameter are defined by the voltage level range that is selected using the niDCPower Configure Voltage Level Range VI.

/images/reference/en-XX/help/370736U-01/cerrcodeclst.gif

error in (no error) describes error conditions that occur before this node runs. This input provides standard error in functionality.

/images/reference/en-XX/help/370736U-01/cbool.gif

status is TRUE (X) if an error occurred before this VI ran or FALSE (checkmark) to indicate a warning or that no error occurred before this VI ran. The default is FALSE.

/images/reference/en-XX/help/370736U-01/ci32.gif

code is the error or warning code. The default is 0. If status is TRUE, code is a non-zero error code. If status is FALSE, code is 0 or a warning code.

/images/reference/en-XX/help/370736U-01/cstr.gif

source describes the origin of the error or warning and is, in most cases, the name of the VI that produced the error or warning. The default is an empty string.

/images/reference/en-XX/help/370736U-01/iio.gif

instrument handle out passes the handle used to identify the session in all subsequent NI-DCPower VI calls.

/images/reference/en-XX/help/370736U-01/ierrcodeclst.gif

error out contains error information. This output provides standard error out functionality.

/images/reference/en-XX/help/370736U-01/ibool.gif

status is TRUE (X) if an error occurred or FALSE (checkmark) to indicate a warning or that no error occurred.

/images/reference/en-XX/help/370736U-01/ii32.gif

code is the error or warning code. If status is TRUE, code is a non-zero error code. If status is FALSE, code is 0 or a warning code.

/images/reference/en-XX/help/370736U-01/istr.gif

source describes the origin of the error or warning and is, in most cases, the name of the VI that produced the error or warning.

WAS THIS ARTICLE HELPFUL?

Not Helpful