niRFSG 5696 Amplifier Cal Initialize VI

NI RF Signal Generators Help (NI-RFSG 18.1)


Edition Date: June 2018
Part Number: 371025V-01
View Product Info

Download Help (Windows Only)


NI-RFSG 18.1 Help
NI-RFSG 18.2.1 Help
NI-RFSG 19.1 Help

Prepares the device for calibrating the amplifier.

Supported Devices: PXIe-5696

niRFSG 5696 Amplifier Cal Initialize

/images/reference/en-XX/help/371025V-01/cio.gif

instrument handle identifies your instrument calibration session. instrument handle is obtained from the niRFSG Initialize External Calibration VI.

/images/reference/en-XX/help/371025V-01/cio.gif

NI 5654 resource name specifies the resource name of the PXIe-5654 RF signal generator to initialize.

/images/reference/en-XX/help/371025V-01/ci32.gif

path to calibrate specifies which amplifier path to calibrate.

High Power Specifies the high power path.
Low Harmonic Specifies the low harmonic path.
/images/reference/en-XX/help/371025V-01/cenum.gif

measurement port specifies the port that the power meter is connected to. To reduce error, the calibration procedure measures the PXIe-5696 input power as well as the PXIe-5696 output power.

Input Power Specifies the PXIe-5696 input power.
Output Power Specifies the PXIe-5696 output power.
/images/reference/en-XX/help/371025V-01/cpath.gif

input power file specifies the file where the input power measurements are stored. This file is written when you set measurement port to Input Power. This file is read when you set measurement port to Output Power.

/images/reference/en-XX/help/371025V-01/cerrcodeclst.gif

error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

/images/reference/en-XX/help/371025V-01/iio.gif

instrument handle out passes a reference to your instrument session to the next VI.

/images/reference/en-XX/help/371025V-01/ierrcodeclst.gif

error out contains error information. This output provides standard error out functionality.