Corrects for any signal drift due to temperature variation when generating the same signal for extended periods of time without a parameter change. Under normal circumstances of short-term signal generation, NI-RFSG performs thermal correction automatically by ensuring stable power levels, and you do not need to call this VI. Use this VI when generating the same signal for a long period of time in a temperature-fluctuating environment. The NI-RFSG device must be in the Generation state before you call this VI.

Supported Devices: PXIe-5611, PXI/PXIe-5650/5651/5652, PXIe-5653/5654/5654 with PXIe-5696, PXI-5670/5671, PXIe-5672/5673/5673E, PXIe-5840

Related Topics

Thermal Management

Impairment Calibration

Examples

niRFSG Perform Thermal Correction

/images/reference/en-XX/help/371025V-01/cio.gif

instrument handle identifies your instrument session. instrument handle is obtained from either the niRFSG Initialize VI or the niRFSG Initialize With Options VI.

/images/reference/en-XX/help/371025V-01/cerrcodeclst.gif

error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

/images/reference/en-XX/help/371025V-01/iio.gif

instrument handle out passes a reference to your instrument session to the next VI. instrument handle out is obtained from the niRFSG Initialize VI or the niRFSG Initialize With Options VI.

/images/reference/en-XX/help/371025V-01/ierrcodeclst.gif

error out contains error information. This output provides standard error out functionality.

Examples

Refer to the RFSG Thermal Correction VI in the labview\examples\instr\niRFSG directory for an example of using the niRFSG Perform Thermal Correction VI.