Short Name: Sync Script Trigger Dist Line

Property of niRFSG

Specifies which external trigger line distributes the synchronized Script Trigger signal. Use the Active Channel property to specify the name of the Script Trigger you are configuring. When synchronizing the Script Trigger, configure all devices to use the same Script Trigger distribution line.

To set this property, the NI-RFSG device must be in the Configuration state.

Supported Devices: PXIe-5644/5645/5646

Default Value: "" (empty string)

Valid Values: PXI_Trig0, PXI_Trig1, PXI_Trig2, PXI_Trig3, PXI_Trig4, PXI_Trig5, PXI_Trig6, PXI_Trig7, PFI0

Related Topics

Script Trigger

Synchronizing Sample Clock and Sampled Reference Clock Signals

Refer to the Synchronization Using NI-RFSA and NI-RFSG topic appropriate to your device for more information about device synchronization for vector signal transceivers.

Remarks

The following table lists the characteristics of this property.

Datatype string
PermissionsRead/Write
High-level VIs N/A
Channel-based Yes
ResettableYes