Device Specific:Vector Signal Transceiver:Triggers:Sync Script Trigger Master Property

NI RF Signal Generators Help (NI-RFSG 18.1)


Edition Date: June 2018
Part Number: 371025V-01
View Product Info

Download Help (Windows Only)


NI-RFSG 18.1 Help
NI-RFSG 18.2.1 Help
NI-RFSG 19.1 Help

Short Name: Sync Script Trigger Master

Property of niRFSG

Specifies whether the device is the master device when synchronizing the Script Trigger between multiple devices. Use the Active Channel property to specify the name of the Script Trigger you are configuring.

The master device distributes the synchronized Script Trigger to all devices in the system through the Script Trigger distribution line.

When synchronizing the Script Trigger, one device must always be designated as the master. The master device actively drives the Script Trigger distribution line. For slave devices, set the Script Trigger Type property to Digital Edge, and set the Script Trigger Digital Edge Source property to sync_script.

To set this property, the NI-RFSG device must be in the Configuration state.

Supported Devices: PXIe-5644/5645/5646

Default Value: FALSE

TRUE The device is the master device for synchronizing the Script Trigger.
FALSE The device is not the master device for synchronizing the Script Trigger.

Related Topics

Script Trigger

Synchronizing Sample Clock and Sampled Reference Clock Signals

Refer to the Synchronization Using NI-RFSA and NI-RFSG topic appropriate to your device for more information about device synchronization for vector signal transceivers.

Remarks

The following table lists the characteristics of this property.

Datatype bool
PermissionsRead/Write
High-level VIs N/A
Channel-based Yes
ResettableYes

WAS THIS ARTICLE HELPFUL?