Converting Voltage Values to Binary Values for the NI 9262/9263/9264/9269 (FPGA Interface)

NI CompactRIO Device Drivers Help

Edition Date: December 2017

Part Number: 373197L-01

»View Product Info
Download Help (Windows Only)

Set the Calibration Mode to Calibrated in the C Series Module Properties dialog box for the NI 9262/NI 9263/NI 9264/NI 9269 if you want the FPGA I/O Node to accept fixed-point data in units of volts when writing to the module. If you set the Calibration Mode to Raw, the FPGA I/O Node accepts only binary values when writing to the module. You must convert output voltage values to binary values before you write them to the module. You must convert these values in the host VI.

Using an Equation to Convert Voltage to Binary

You can use the following equation in the host VI to convert the analog output values to binary values:

Binary Value = (Voltage Value × 109Offset) ÷ LSB Weight,

where Binary Value is the value you write to the FPGA I/O Node
Voltage Value is the voltage in V that you want the channel to output
Offset is the value returned by the Offset property
LSB Weight is the value returned by the LSB Weight property.

NI recommends using calibrated values for analog output. To convert calibrated voltage values, use the FPGA I/O Property Node to read the LSB Weight and Offset properties. If you do not want to read the LSB Weight and Offset values from the module, you can convert uncalibrated voltage values by using the following values for Offset and LSB Weight:

Offset = 0

LSB Weight = Typical Output Span ÷ 2DAC Resolution × 109

where Typical Output Span is 21.4 V for the NI 9263, 21 V for the NI 9264, and 20.98 V for the NI 9269
DAC Resolution is the DAC resolution value in the hardware documentation for the module.