Scope Run VI

LabVIEW 2018 ELVIS III Toolkit Help

Edition Date: March 2018

Part Number: 377208A-01

»View Product Info
Download Help (Windows Only)

Owning Palette: Oscilloscope VIs

Requires: ELVIS III Toolkit

Transitions the acquisition from the Stopped state to the Running state. If the current state is Triggered, the acquisition is first transitioned to the Stopped state before transitioning to the Running state.

Instrument Handle In specifies the session created by the Scope Initialize VI.
Auto Trigger specifies whether to trigger the instrument automatically if the trigger condition you configure is not met within the timeout period.
TRUEThe instrument is triggered automatically if the configured trigger condition is not met within the timeout period. When pretrigger time is equal to or greater than 0.25 second, the timeout period is the trigger delay plus pretrigger time * 2. If pretrigger time is less than 0.25 second, the timeout period is the trigger delay plus 0.5 second.
FALSEThe instrument waits indefinitely for any occurrence of trigger condition.


The default value is FALSE.
Error In describes error conditions that occur before this node runs. This input provides standard error in functionality.
Instrument Handle Out returns a reference to the instrument.
Error Out contains error information. This output provides standard error out functionality.