Features Supported on PXIe-5413/5423/5433 Waveform Generators

NI Signal Generators Help (NI-FGEN 18.1)

Edition Date: April 2018

Part Number: 370524V-01

»View Product Info
Download Help (Windows Only)
PXIe-5413 PXIe-5423 PXIe-5433
Basic Operation
Output Modes Standard Function, Frequency List, Arbitrary Waveform, Arbitrary Sequence Standard Function, Frequency List, Arbitrary Waveform, Arbitrary Sequence, Script5 Standard Function, Frequency List, Arbitrary Waveform, Arbitrary Sequence, Script5
Independent Channels Two-channel PXIe-5413 only6 Two-channel PXIe-5423 only6 Two-channel PXIe-5433 only6
Standard Function Output
Waveform Sine, Square, Triangle, Ramp Up, Ramp Down, DC, Noise, User-Defined Sine, Square, Triangle, Ramp Up, Ramp Down, DC, Noise, User-Defined Sine, Square, Triangle, Ramp Up, Ramp Down, DC, Noise, User-Defined
Minimum Frequency 0 Hz 0 Hz 0 Hz
Maximum Frequency 3 Sine: 20 MHz, Square: 10 MHz, Ramp and Triangle: 1 MHz, User: 20 MHz, Other: 20 MHz Sine: 40 MHz, Square: 25 MHz, Ramp and Triangle: 5 MHz, User: 40 MHz, Other: 40 MHz Sine: 80 MHz, Square: 50 MHz, Ramp and Triangle: 50 MHz, User: 80 MHz, Other: 80 MHz
SYNC Duty Cycle
User-Defined Waveform Size 8,192 samples 8,192 samples 8,192 samples
Frequency List Output
Maximum Number of Lists* 9,999 9,999 9,999
Maximum List Length * 1,024 1,024 1,024
Maximum Step Duration * 2,814,749 s 2,814,749 s 2,814,749 s
Minimum List Length * 1 1 1
Minimum Step Duration * 10 ns 10 ns 10 ns
Step Duration Quantum * 10 ns 10 ns 10 ns
Arbitrary Waveform Output
Write Quantum 1 sample 1 sample 1 sample
Waveform Quantum * 1 sample 1 sample 1 sample
Minimum Waveform Size * 2 samples 2 samples 4 samples
Maximum Waveform Size * 1 64 M samples 64 M samples 256 M samples
Maximum Number of Waveforms *  1,048,572  1,048,572 4,194,304
Streaming Yes Yes
Onboard Signal Processing
Arbitrary Sequence Output
Minimum Sequence Length * 1 1 1
Maximum Sequence Length * 65,535 65,535 65,535
Maximum Loop Count * 16,777,215 16,777,215 16,777,215
Maximum Number of Sequences * 65,535 65,535 65,535
Streaming Yes Yes
Onboard Signal Processing
Script Output
Maximum Number of Script Triggers 4 4
Maximum Number of Markers 4 4
Streaming Yes Yes
Output Characteristics
Output Voltage (at load equal to source impedance) up to ±6 V up to ±6 V up to ±6 V
Offset (at maximum gain) ±3 V ±3 V ±3 V
Output Impedance 50 Ω 50 Ω 50 Ω
Analog Path Low-Gain, High-Gain Low-Gain, High-Gain Low-Gain, High-Gain
Analog Filter Option
Filter Correction Frequency Option Yes Yes Yes
Digital Filter Option Yes Yes Yes
Digital Filter Interpolation Factor
DIGITAL DATA & CONTROL CONNECTOR (DDC) or Digital Pattern
Triggering and Synchronization
Trigger Modes (Frequency List and Arbitrary Waveform Generation Modes) Single, Continuous, Stepped, Burst Single, Continuous, Stepped, Burst Single, Continuous, Stepped, Burst
Trigger Sources Immediate, Software, PXI_TRIG <0..7>, PFI <0..1>, AUX 0/PFI <0..7> Immediate, Software, PXI_TRIG <0..7>, PFI <0..1>, AUX 0/PFI <0..7> Immediate, Software, PXI_TRIG <0..7>, PFI <0..1>, AUX 0/PFI <0..7>
Multiple Device Synchronization Using NI-TClk 6 (NI-FGEN 18.1 or later) Using NI-TClk 6 (NI-FGEN 18.1 or later) Using NI-TClk 6 (NI-FGEN 18.1 or later)
Multiple Channel Synchronization Automatic in the same NI-FGEN session Automatic in the same NI-FGEN session Automatic in the same NI-FGEN session
Distributed TClk Yes 7 Yes 7 Yes 7
Events
Ready for Start Yes Yes Yes
Started Yes Yes Yes
Done Yes Yes Yes
Marker Yes Yes Yes
Data Marker
Clocking
Sample Rate (Update Rate) before filtering and interpolation 3.125 MS/s to 200 MS/s 3.125 MS/s to 200 MS/s 10 S/s to 400 MS/s with digital filter enabled, 250 MS/s with digital filter disabled
Reference Clock Source Internal (none), PXI Clock (PXIe_Clk100) Internal (none), PXI Clock (PXIe_Clk100) Internal (none), PXI Clock (PXIe_Clk100)
Reference Clock Frequency 100 MHz 100 MHz 100 MHz
Clock Mode (Arbitrary Waveform Generation Mode) High-Resolution High-Resolution High-Resolution
Sample Clock Source (Arbitrary Waveform Generation Mode) Internal Internal Internal
Calibration
Self-Calibration Functions niFgen_SelfCal niFgen_SelfCal niFgen_SelfCal
Calibration Utility Functions niFgen_ functions 4 niFgen_ functions 4 niFgen_ functions 4
External Calibration Functions 2 ni5433_ functions ni5433_ functions ni5433_ functions

*You can get the value of this characteristic by calling a query function or by reading an attribute. NI recommends that your programs query or read the characteristic rather than depend on a certain value.

1Varies by waveform generator and the amount of memory. Memory use is a function of the number and size of waveforms and (in Arbitrary Sequence mode) the number and length of sequences. Typically, waveforms use most of the memory, but if you have a very large number of sequences, the available waveform memory is reduced.

2External calibration functions and steps vary by waveform generator. Refer to the calibration procedure for your waveform generator for more information.

3Refer to the specifications for your waveform generator for conditions.

4The niFgen Get Ext Cal Last Temp VI and niFgen Restore Last Ext Cal Constants VI are not supported.

5if/else/end if and break script instructions are not supported.

6Synchronization with NI-TClk is supported with a single NI-FGEN session per waveform generator. Synchronization with NI-TClk is not supported with multiple channel-based NI-FGEN sessions to a single multichannel waveform generator. Refer to the Synchronization topic for your PXIe-5413/5423/5433 waveform generator for more information.

7Refer to the Distributed TClk topic in the NI-TClk Synchronization Help for information about distributed TClk.

WAS THIS ARTICLE HELPFUL?