Voltage High

NI High-Speed Digitizers Help (NI-SCOPE)

Edition Date: January 2017

Part Number: 370592AB-01

View Product Info

DOWNLOAD (Windows Only)


NI-SCOPE (English | Japanese)

C/C++ Equivalent: NISCOPE_VAL_VOLTAGE_HIGH

The last-acquisition histogram method is used, where the voltage high result is the voltage of the histogram bin with the maximum number of hits above 60% of the waveform's voltage peak-to-peak value. This calculation is useful for ignoring the overshoot and preshoot on square waves.