NISCOPE_ATTR_TRIGGER_WINDOW_HIGH_LEVEL

NI High-Speed Digitizers Help (NI-SCOPE)

Edition Date: January 2017

Part Number: 370592AB-01

View Product Info

DOWNLOAD (Windows Only)


NI-SCOPE (English | Japanese)

Specific Attribute

Data
type
Access Applies to Coercion High Level Functions
ViReal64 R/W N/A None niScope_ConfigureTriggerWindow

Description

The upper voltage threshold you want the digitizer to use for window triggering.

The digitizer triggers when the trigger signal enters or leaves the window you specify with NISCOPE_ATTR_TRIGGER_WINDOW_LOW_LEVEL and this attribute.

Valid Values: The value you pass for this parameter must meet the following conditions:

High Trigger Level <= Vertical Range/2 + Vertical Offset

High Trigger Level >= (–Vertical Range/2) + Vertical Offset

High Trigger Level > Low Trigger Level