NISCOPE_ATTR_TRIGGER_WINDOW_LOW_LEVEL

NI High-Speed Digitizers Help (NI-SCOPE)

Edition Date: January 2017

Part Number: 370592AB-01

View Product Info

DOWNLOAD (Windows Only)


NI-SCOPE (English | Japanese)

Specific Attribute

Data
type
Access Applies to Coercion High Level Functions
ViReal64 R/W N/A None niScope_ConfigureTriggerWindow

Description

The lower voltage threshold you want the digitizer to use for window triggering.

The digitizer triggers when the trigger signal enters or leaves the window you specify with this attribute and NISCOPE_ATTR_TRIGGER_WINDOW_HIGH_LEVEL.

Valid Values: The value you pass for this attribute must meet the following conditions:

Low Trigger Level <= Vertical Range/2 + Vertical Offset

Low Trigger Level >= (–Vertical Range/2) + Vertical Offset

Low Trigger Level < High Trigger Level