Onboard Signal Processing:OSP Overflow Error Reporting Property

NI High-Speed Digitizers Help (NI-SCOPE)

Edition Date: January 2017

Part Number: 370592AB-01

View Product Info

DOWNLOAD (Windows Only)


NI-SCOPE (English | Japanese)

Short Name: Overflow Error Reporting

Property of niScope

Configures error reporting when the onboard signal processing block detects an overflow in any of its stages. Overflows lead to clipping of the waveform. The default value is Warning.

Error (0)

Execution stops and NI-SCOPE returns an error when an overflow has occurred in the OSP block.

Warning (1)

Execution continues and NI-SCOPE returns a warning when an overflow has occurred in the OSP block.

Disabled (2)

NI-SCOPE does not return an error when an overflow has occurred in the OSP block.

Remarks

The following table lists the characteristics of this property.

Datatype iL
PermissionsRead/Write
High-level VIs N/A
Channel-based No
ResettableNo

WAS THIS ARTICLE HELPFUL?

Not Helpful