Short Name: Trigger Window Low Level
Property of niScope
Pass the lower voltage threshold you want the digitizer to use for window triggering.
The digitizer triggers when the trigger signal enters or leaves the window you specify with this property and the Trigger Window High Level property.
The values of the Vertical Range property and the Vertical Offset property determine the valid range for this property on the channel you specify with the Trigger Source property.
The value you pass for this parameter must meet the following conditions:
Low Trigger Level <= Vertical Range/2 + Vertical Offset
Low Trigger Level >= (-Vertical Range/2) + Vertical Offset
Low Trigger Level < High Trigger Level
Related topics:
The following table lists the characteristics of this property.
Datatype | ![]() |
Permissions | Read/Write |
High-level VIs | niScope Configure Trigger Window (Analog Window Ref Trigger) |
Channel-based | No |
Resettable | No |
Helpful