niScope Cal End

NI High-Speed Digitizers Help (NI-SCOPE)

Edition Date: January 2017

Part Number: 370592AB-01

View Product Info

DOWNLOAD (Windows Only)


NI-SCOPE (English | Japanese)

Closes an external calibration session.

If action is set to Abort Calibration, the session is closed and the new calibration constants are lost. Some devices may write to the EEPROM during calibration, in which case the Abort Calibration action restores the EEPROM to its original state. It is, therefore, very important to call this VI each time you call niScope Cal Start, even if an error occurs during calibration.

If action is set to Store Calibration, the new calibration constants are stored in the EEPROM. For most digitizers, the current system date and the incremented calibration count are also stored; for SMC-based digitizers, the current system date and onboard temperature are stored.

niScope Cal End

/images/reference/en-XX/help/370592AB-01/cio.gif

instrument handle is the instrument handle that you obtain from niScope Cal Start. The handle identifies a particular instrument session.

/images/reference/en-XX/help/370592AB-01/ci32.gif

action either stores the calibration constants or aborts the calibration and discards any calibration results.

Defined Values

Store Calibration

Abort Calibration

/images/reference/en-XX/help/370592AB-01/cerrcodeclst.gif

error in describes error conditions that occur before this node runs. This input provides standard error in functionality.

/images/reference/en-XX/help/370592AB-01/ierrcodeclst.gif

error out contains error information. This output provides standard error out functionality.