NI-DCPower (English | Japanese) |
Short Name: Output Enabled
Property of niDCPower
Specifies whether the output is enabled (TRUE) or disabled (FALSE).
Depending on the value you specify for the Output Function property, you also must set the voltage level or current level in addition to enabling the output.
This property has no effect on the output disconnect relay. To toggle the relay, use the Output Connected property.
![]() |
Note If the session is in the Committed or Uncommitted states, enabling the output does not take effect until you call the niDCPower Initiate VI. Refer to the Programming States topic in the NI DC Power Supplies and SMUs Help for more information about NI-DCPower programming states. |
Default Value: Refer to Supported Properties by Device for the default value by device.
The following table lists the characteristics of this property.
Datatype | ![]() |
Permissions | Read/Write |
High-level VIs | niDCPower Configure Output Enabled |
Channel-based | Yes |
Resettable | No |
Helpful