This topic contains information for calibrating the NI PXIe-4339 using Calibration Executive.
Calibration Executive Procedure Features:
Adjustment![]() |
Manual Mode![]() |
Selectable Test Points![]() |
Approximate Test Time:
Verify Only 23 minutes |
Verify & Adjust 43 minutes |
Test Equipment | | | Test Conditions | | | Device Setup | | | Test Limit Equations |
The following table lists the test equipment required to calibrate the PXIe-4339.
Instrument | Recommended Model | Requirements |
---|---|---|
Calibrator | Fluke 5700 | If this instrument is unavailable, use a calibrator that can provide DC voltage values in the range of 0 to ±10 V with an accuracy of 50 ppm of output or better and a DC offset error of 4 µV or better. |
PXI Express Chassis | PXIe-1062Q | — |
Connection Accessory | TB-4339/B/C* | — |
* The terminal block variants TB-4339, TB-4339B, and TB-4339C differ only in resistance; the measurements are the same regardless of the variant used. |
The following setup and environmental conditions are required to ensure the PXIe-4339 meets published specifications.
The following test limits are derived from the published specifications.
Voltage Mode
Ratiometric Mode
Refer to the NI PXIe-4339 Device Specifications for Gain Error and Offset Error values.