Calibration

RFmx WCDMA 19.1 Help

Edition Date: October 2019

Part Number: 375421H-01

»View Product Info

DOWNLOAD (Windows Only)


RFmx WCDMA 19.1 Help
RFmx WCDMA 20.0 Help
RFmx WCDMA 20.6 Help
RFmx WCDMA 20.7 Help
Owning Palette: RFmxInstr VIs

Use the VIs on this palette to configure calibartion measurements. You can use the RFmxInstr Property Node to configure additional properties.

Palette Object Description
RFmxInstr Self Calibrate

Self-calibrates the NI-RFSA device and associated modules that support self-calibration. If self-calibration completes successfully, the new calibration constants are stored immediately in the nonvolatile memory of the module.

Refer to the specifications document for your device for more information about how often to self-calibrate. For more information about Self Calibrate, refer to the niRFSA Self Cal VI topic for your device in the NI RF Vector Signal Analyzers Help.

RFmxInstr Self Calibrate Range

Self-calibrates all configurations within the specified frequency and reference level limits. If there is an open session for NI-RFSG for your device, it may remain open but cannot be used while this VI runs. NI recommends that no external signals are present on the RF In port while the calibration is taking place. For more information about Self Calibrate Range, refer to the niRFSA Self Calibrate Range VI topic for your device in the NI RF Vector Signal Analyzers Help.

RFmxInstr Is Self Calibrate Valid

Returns an array to indicate which calibration steps contain valid calibration data. To omit steps with the valid calibration data from self-calibration, you can pass the Valid Steps parameter to the Steps To Omit parameter of the RFmxInstr Self Calibrate VI.



Subpalette Description
Calibration Utility

Use the VIs on this palette to configure calibration utility measurements. You can use the RFmxInstr Property Node to configure additional properties.